Total cycle time: Standard cycle, 28 minutes; Advanced cycle, 38 minutes

Cycle temperature: Less than 131°F (55°C)

External dimensions:
22 in (56 cm) wide
33 in (84 cm) high
33 in (84 cm) deep

Weight:
Sterilizer–275 lb (125 kg)
Cart–100 lb (45 kg)

Sterilization chamber:
Rectangular configuration 1.1 ft3 (30 liters)
Two-tiered shelf–12.3 in x 23.6 in (31.2 cm x 60 cm)

Installation and Operating Requirements

Connections:
12 AWG, 98 in (249 cm) long
NEMA 5-20P; 108-132 VAC/ 60 Hz/ 16 A

Placement: Tabletop, or cart-mounted for mobility

Operational environment:
Temperature–64°F to 95°F (18°C–35°C)
Humidity–10% to 85% up to 86°F, linearly decreasing from 85% at 86°F to 70% at 95°F

*Regulatory approvals:
- CAN/CSA-C22.2 No. 1010.1-92; Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use
- UL 3101-1; Standard for Safety for Electrical Equipment for Laboratory Use
- EN 61010; Safety Requirements for Electrical Equipment for Measurement, Control, and Laboratory Use.
- EN 55011; Group II Class A limits, based on CISPR 11; Emissions
- FCC 47 Part 18: FCC Requirements for Industrial, Scientific, and Medical Equipment
- IEC 601-1-2; Medical Electrical Equipment, Collateral Standard, Electromagnetic Compatibility
- CE, a mark of conformity for the European Community

STERRAD NX Sterilization System